Print menu, Parts counting (cont.) – Ohaus Scout II Electronic Balances User Manual
Page 11
11
PARTS COUNTING (CONT.)
7. To read the weight of the parts, press Mode Off to change to any of the
activated weighing units.
8. To return to a weighing mode, press Mode Off to select a weighing
unit, the display indicates the weight of the container as a negative value.
Simply press Zero On to tare the balance. Taring the balance does not
affect the sample weight which is stored and retained as long as the
balance remains on.
9.
To count similiar parts,
place a container on the pan and press Zero On, then place parts in the
container. The balance displays the number of parts.
PRINT MENU
The Print menu is used to configure and customize the RS232 interface
parameters for your requirements on balances equipped with this function.
The following table shows the sequence in which submenus appear on the
Print Menu. Factory settings are in bold.
PRINT MENU TABLE
POWER When set ON, enables the operation of the RS232. Not
recommended for battery operation.
Baud Rate Specifies baud rate of either 300, 1200, 2400, 4800, or 9600.
Data Bits Specifies number of data bits, 7 or 8.
Parity Bit Specifies parity type, Odd, Even, or None.
Stop Bits
Specifies the number of stop bits, 1 or 2.
Auto Print
Enables either continuous, on stabilty, off or automatic
printing.
Stable Data
Enables/disables printing stable data only feature.
End
Used to exit the Print menu and store your selections.