4220 Flow Meter

Section 6 Maintenance and Service

6-11

Note

If you disassemble the flow meter for servicing, you will also
remove the aluminum chassis covers to access the circuitry.
Always replace these covers when repairs have been com-
pleted. The covers protect the circuit boards and also reduce
signal emissions that could interfere with the operation of
nearby electronic equipment. For the same reason, do not
remove any of the ferrite beads or alter the wiring harnesses
inside the cabinet in any way.

Figure 6-6 Lifting the Flow Meter from the Case

6.5.2 Fuse Replacement

With the flow meter chassis out of the cabinet, you can locate and
change fuses. The fuses are located on the printed circuit board
directly behind the keypad. There is an aluminum cover over this
board. Remove the cover by pressing against its surface with the
palm of your hand and then pulling downward on it with your
fingers. A plastic catch under the cover holds it in place. The
fuses are labeled F1, F2, and F3. The proper size for each of these
fuses is:

F1 - 5 amp., fast blow

F2 - 5 amp., fast blow

F3 - 2

1

/

2

amp., fast blow

Always replace a blown fuse with one of the same value. Using a
larger value fuse could cause serious damage to the flow meter or
to its power supply. Replace the protective cover, making sure the
two ears on top of the cover slide into the mating slots on the
chassis. There should be a noticeable snap when the plastic catch