5.13 Mechanical Totalizer

A mechanical totalizer is available for the 4220 that consists of a
seven-digit, non-resettable mechanical counter mounted in the
front panel. It must be ordered with the flow meter. The totalizer
advances according to program selections for units of measure
and the maximum flow of the primary device used. The totalizer
is internally set to advance at

1

100

of the rate of the display

totalizer. Consequently, you must multiply the number shown on
the mechanical totalizer by 100 to determine the actual value for
total flow.

Accuracy

± 0.4° C

Resolution 0.1°

C

Dissolved Oxygen % Saturation

Sensor Type

Rapid Pulse - Polarographic

Range

0 to 200 % Air Saturation

Accuracy

± 2% Air Saturation

Resolution

0.1% Air Saturation

Dissolved Oxygen mg/L

Sensor Type

Calculated from % air saturation, temperature,
and salinity

Range

0 to 20 mg./L

Accuracy

± 0.2 mg./L

Resolution 0.01

mg./L

pH

Sensor Type

Glass Combination Electrode

Range

2 to 14 pH Units

Accuracy

± 0.2 Units

Resolution

0.01 Units

Conductivity*

Sensor Type

Four-Electrode Cell

Range

0 to 100 mS/cm

Accuracy

± (1% of reading + 0.001 mS/cm)

Resolution

0.01 mS/cm or 1

μS/cm

*Specific conductance (conductivity corrected to 25° C), resistivity, and total dis-
solved solids measurements are also provided. These values are automatically calcu-
lated from conductivity according to algorithms found in Standard Methods for the
Examination of Water and Wastewater
(ed. 1989).

Salinity

Sensor Type

Calculated from conductivity and temperature

Range

0 to 70 ppt

Accuracy

± 0.2 ppt

Resolution

0.1 ppt

Table 5-5 YSI 600 Probe Specifications (Continued)