Troubleshooting Guide

6

6-2

Problem

Solution

The thermometer is

Remove and reinstall the batteries or

“locked up” (the

disconnect and reconnect the ac

display is “frozen”).

adapter.

The display is either

1.

Clean the thermometer lens.

erratic or stays at one

Refer to Section 4.2.

reading.

2.

Activate the Diagnostic Program in
the thermometer as follows:

a. Pull the trigger and press the

key to lock the trigger.

b. Hold down the

key and

press the

key until

“VER X.X” is displayed.

You can expect to see and hear the
following:

You will see the version number
“VER X.X” of the software for about
1 second.

You will hear a beep, “TST” is
displayed and °F flashes.

Soon after, all of the segments of
the display including the
backlighting will light up for about
1 second.

The display will clear and a PAS
(pass) or ERR (error) code may be
seen on the display.