Using the Handheld Infrared Thermometer

2

2-18

8.

Press and hold the

key until the Emissivity Display

Mode (E) appears.

9.

Change the emissivity to the proper value for the target
being measured (refer to Section 2.3.1).

10. Aim at the target. The target temperature and emissivity

are displayed on the LCD.

11. After all data is taken, press the

key to release

this mode

To disable the Ambient Target Temperature

Compensation at a later time, you must press the

key to get into the Ambient Target

Temperature Display Mode. Then press the

key to disable it.

The target ambient temperature does not change
when the thermometer is turned off. However,
when the batteries are replaced, it is reset to the
default value of 75°F.

2.3.9 Sending Temperature Data to a Serial Printer

(OS533, OS534, OS523, OS524)

The thermometer can transmit temperature data to a Serial
Printer via the RS-232 phone jack and the RS-232 cable.

1.

Turn on the Serial Printer and set the following

parameters:

Speed: 4800 BPS

One Stop Bit

Data:

8 Bits

No Parity

2.

Connect the RS-232 cable between the thermometer and
the printer as shown in Figure 2-14.