11.6 Device Control 

11.6.1 Output for the Target Temperature 

The signal output can be set to 4 – 20 mA, 0 – 20 mA or mV. If current 
output is activated, the output can provide a predefined current: 

XO=4 

output mode to 4 – 20 mA 

O=13.57 

output of a constant current at 13.57 mA 

O=60 

switches back to the temperature controlled output 

11.6.2 Analog Output, Scaling 

According to the temperature range of the model, it is possible to set 
a  maximum  voltage/current  value  according  to  a  temperature  value 
(e.g., the maximum current 20 mA shall represent 200°C / 392°F). The 
same setting is possible for the minimum value. 

H=500 

the maximum current/voltage value is set to 500°C 

L=0 

the minimum current/voltage value is set to 0°C 

Remark:  You  cannot  set  this  value  for  thermocouple  output.  The 
minimum span between the maximum / minimum settings is 20 K. 

11.6.3 Alarm Output 

The second output channel can be set in different modes, see section 
5.3.2 

Head Ambient Temp. / Alarm Output

 on page 20. 

• Internal sensing head temperature 
• Alarm output 

K=7 

internal sensing head temperature 

K=4 

alarm  output  for  object  temperature,  0  V  in  case  of  no 
alarm 

XS=125.3  threshold setting to 125.3°C (if U=C is set) 

11.6.4 Factory default values 

It is possible to reset the unit to the default values.