beautypg.com

5 electrical interference, Basics mi 13 4.5 electrical interference – RayTek MI Miniature Infrared Sensor User Manual

Page 21

background image

Basics 

MI 

13 

4.5 Electrical Interference 

To  minimize  electrical  or  electromagnetic  interference  or  “noise”  be 
aware of the following: 

• Mount the unit as far away as possible from potential sources 

of  electrical  interference  such  as  motorized  equipment 
producing large step load changes. 

• Use shielded wire for all input and output connections. 
• Make sure the shield wires are earth grounded at one point. 
• Sensor  head  shield  braid  should  make  direct  contact  around 

the cable circumference.