2 wiring, 1 sensor head cable, Installation mi 15 – RayTek MI Miniature Infrared Sensor User Manual
Page 23
Installation
MI
15
Figure 4: Proper Sensor Placement
5.2 Wiring
5.2.1 Sensor Head Cable
The manufacturer preinstall’s the sensor head cable between sensor
head and electronic box. It may be shortened but not lengthened.
Shortening the cable length by 1 m (3 ft.) causes a
temperature error of – 0.1 K / m!
Do not bend the sensor head cable tighter than 25 mm / 1 in.
(MID/MIC) and 15 mm / 0.6 in. (MIH) respectively!
Target greater than spot size
Target equal to spot size
Target smaller than spot size
best
good incorrect
Sensor