Programming guide mi 71 – RayTek MI Miniature Infrared Sensor User Manual
Page 79

Programming Guide
MI
71
There are eight entries possible for emissivity setting (1) and a related
set point (threshold) (2). To be able to write or read these values, use
the following commands:
EP=2
set pointer for table entry, e.g. to line 2 (3)
RV=0.600
set the emissivity value for line 2 to 0.600 (4)
SV=220.0
set the set point (threshold) for line 2 to 220.0 (5)
Figure 46: Table for Emissivity and Set Points
To activate these emissivity settings, you need to have the 3 external
inputs (FTC) connected. According to the digital combination on the
FTC wires, one of the table entries will be activated, see section 5.4.2
Emissivity Setting via Digital Selection
1
2
3
4
5