beautypg.com

Triggers analog trigger, Digital trigger, Rtsi – National Instruments AT E Series User Manual

Page 132

background image

Appendix A

Specifications for AT-MIO-16XE-10 and AT-AI-16XE-10

© National Instruments Corporation

A-27

AT E Series User Manual

Triggers

Analog Trigger

Source..................................................... ACH<0..15>, PFI0/TRIG1

Level....................................................... ± Full-scale, internal;

±10 V, external

Slope....................................................... Positive or negative

(software selectable)

Resolution .............................................. 12 bits, 1 in 4,096

Hysteresis ............................................... Programmable

Bandwidth (–3 dB)................................. 255 kHz internal,

4 MHz external

External input (PFI0/TRIG1)

Impedance ....................................... 10 k

Coupling.......................................... DC

Protection ........................................ –0.5 to Vcc +0.5 V when

configured as a digital signal

±35 V when configured as an
analog trigger signal or disabled

±35 V powered off

Accuracy ................................................ ±1% of full-scale range

Digital Trigger

Compatibility ......................................... TTL

Response ................................................ Rising or falling edge

Pulse width ............................................. 10 ns min

RTSI

Trigger Lines.......................................... 7