Dynamic characteristics, Trigger characteristics, Stability – National Instruments SCXI-1127 User Manual
Page 70: Physical, Environment, Safety
Appendix A
Specifications for SCXI-1128
© National Instruments Corporation
A-5
Dynamic Characteristics
Maximum operating speed..................... 1200 cycles/s
Relay operate time (at 20
°C) ................ 0.25 ms typical, 0.5 ms max
Relay release time (at 20
°C)................. 0.08 ms typical, 0.2 ms max
Trigger Characteristics
Scanner advanced pulsewidth ................ 1.1
µs
External trigger input pulsewidth........... 500 ns (min)
Stability
Recommended warm-up time ................ 5 minutes
Physical
Dimensions............................................. 3.0 by 17.2 by 20.3 cm
(1.2 by 6.9 by 8.0 in.)
Environment
Operating temperature ..........................0 to 50
°C
Storage temperature..............................–20 to 70
°C
Relative humidity ................................... 10% to 90% noncondensing
Electromagnetic compatibility (EMC)
emissions ................................................ Refer to the document supplied
with the chassis for compliance to
relevant directives.
EMI ........................................................ FCC Class A compliant
Safety
Designed in accordance with IEC-61010-1, UL 3111-1, and
CAN/CSA C22.2 No. 1010.1 for electrical measuring and test equipment
Approved at altitudes up to 2000 m.
Installation Category II