Command Register 2

(Continued)

Bit

Name

Description

0

CALEN

Calibration Enable Bit—If this bit is set, the auto-
calibration of the 12 bit ADC is enabled. The power-
on value is 0. To start the auto-calibration, first write
one to this bit, then read this register. The result of the
reading is ignored. An auto-calibration lasts about
10 ms. By checking the CONVPROG bit of the Status
Register, the completion of auto-calibration can be
detected. After the auto-calibration, you must clear
this bit for the A/D conversion operation.

a.Book : l.Appendix D Page 8 Wednesday, November 20, 1996 6:36 PM