Sierra Monitor Corporation 5100-06-IT User Manual
Page 45

Model 5100-XX-IT Toxic Gas Sensor Module
Page: 41
Command 132: Set Warning Level
Request Data Bytes:
Byte
Format
Description
0
Unsigned-8
Warning level, range 0 to 60. Default is 20.
Command Specific Response Data Bytes:
Byte
Format
Description
0
Unsigned-8
Returns new Warning Level.
Command 133: Set Alarm Relay Action
Request Data Bytes:
Byte
Format
Description
0
Unsigned-8
0=Latching (default)
2=Non-Latching
Command Specific Response Data Bytes:
Byte
Format
Description
0
Unsigned-8
Returns new Alarm Relay Action value
Command 134: Set Warning Relay Action
Request Data Bytes:
Byte
Format
Description
0
Unsigned-8 0=Latching
2=Non-Latching (default)
Command Specific Response Data Bytes:
Byte
Format
Description
0
Unsigned-8
Returns new Warning Relay Action value
Command 135: Reset Alarms
This command will reset any alarm relays that have been latched, providing the alarm condition is no longer
present on the sensor.
Request Data Bytes:
None
Command Specific Response Data Bytes: None
Command 136: Abort Calibration
This command aborts the calibration procedure.
Request
Data
Bytes:
None
Command Specific Response Data Bytes: None
Command 137: Set Calibration Gas Level
Request Data Bytes:
Byte
Format
Description
0
Unsigned-8
Calibration Gas Level, range 0 to 100
Command Specific Response Data Bytes:
Byte
Format
Description
0
Unsigned-8
Returns new Calibration Gas Level