MediSense PCx User Manual

Page 77

background image

Questions? Refer to Page 10-13 For Customer Care Information.

8-17

15. The administrator will

select one of the
following options:

Press 1

Disabled.

Press 2

Optional.

Press 3

Required.

The display is returned to
the Patient Test Comment
Code If In-Range screen.

16. Press 1

Next;

continue to the next
menu.

The display shows the
Out-of-Range Comment
Code menu. The comment
code options can be set to
Disabled, Optional or
Required for test results
that fall outside the action
range.

What You See
on the Display

What You Do

Comments

17. The administrator will

select one of the
following options:
Press 1

Next, and

continue to step 20.
Press 2

Change, to

change the Out-of-
Range Comment
Code.

The display shows the
Out-of-Range Comment
Code options.

18. The administrator will

select one of the
following options:

Press 1

Disabled.

Press 2

Optional.

Press 3

Required.

The display shows the
Patient Test Comment
Code If Out-of-Range
screen.