4-5

14. The operator may

select one of the
following:

Press 1 – Next Level.

Press 2 – Repeat Test

Press Menu to return to
the Menu Mode menu.

Press On/Off to turn off
the monitor.

If 1 is selected, the next
level test will appear. The
operator may perform the
next level control test by
returning to Step 4.

If 2 is selected, the Scan
or Enter Strip Lot screen
will appear.

Return to step 5 to repeat
the test.

What You See
on the Display

What You Do

Comments

13. Remove the test strip

from the monitor
when finished testing.

(2.7 mmol/L)

11. Note the test result

and whether it falls
within the acceptable
range.

The display shows the
control level, PASS/FAIL,
date and time.

12. If required, scan or

manually enter the
comment code, and
press Enter.

The monitor may be
enabled to scan or
enter a 1- to 2-digit
comment code. If
there is no prompt to
enter a comment code,
skip down to step 13.

An “X” will appear in
front of a test result that is
Out-of-Range.

(x3.7 mmol/L)