Control test procedure – MediSense PCx User Manual

Page 24

background image

4-2

What You See
on the Display

What You Do

Comments

Control Test Procedure

Use the following procedure to perform control tests.

Note: The word “Strip” on the display always refers to the test
strip. If the monitor displays a message not indicated in these
procedures, please refer to section 10, Troubleshooting.

1. Press On/Off to turn on

the monitor.

The Abbott logo screen
will appear for a few
seconds followed by the
product name, software
version and then the Test
Mode menu.

2. Press 2 – Control Test.

After the control test is
selected, the Operator ID
screen may appear, if the
setting has been enabled.

3. Press Scan to scan the

Operator ID barcode, or
manually enter the
Operator ID via the
keypad, then press
Enter.

If the Operator ID is not
required, just press enter.

After the Operator ID is
entered, the Low Level
Solution Lot screen will
appear.

4. Scan or manually enter

the low control solution
lot number via the
keypad, then press
Enter.

The Strip Lot screen will
appear next.

If the Unexpected Level
screen appears, the
operator may either:

1. Enter 1 to ReEnter the

expected level.

2. Enter 2 to continue to

test the unexpected
level.