Symptom/Error
Message

Explanation

How To Respond

The test was not
successful. An error was
detected during the
analysis sequence.

1. Remove the old test

strip.

2. Insert a new test strip.

If the error occurs
again, record the three-
digit error code and
contact MediSense
Customer Care.

Press 2 to exit to the Test
Menu.

The test strip
malfunctioned or the
blood glucose level is
beyond the measuring
capabilities of the system.

1. Remove the used test

strip.

2. Insert a new test strip.

3. Press 1 to continue

testing with a new test
strip.

4. If the error appears

again, confirm the
result by performing a
laboratory reference
test. Contact
MediSense Customer
Care.

(Error codes 112,

114, 116-118)

(Error codes 113, 115)

If the Scan button is
pressed and the laser has
been on for a total of 60
seconds in the last 15
minutes, the monitor will
beep and display Laser
on Time Exceeded
.

Turn the monitor off and
wait 15 minutes for the
laser time to reset.

Note: Remember to
release the Scan button
before trying to rescan.

During the test, battery
power was too low for
testing.

Do one of the following:

• Install new batteries. See

Maintenance for more
information.

• Place the monitor in the

docking station (if the
system has this option).

Press 1 to exit.