Maintenance

6

Chaning the Syringe Piston

Assembly

Changing the Syringe Piston
Assembly

For all syringe sizes, it is necessary to replace the

piston assembly after approximately 30 000 cycles

(using water).

The correct reference piston kit must be ordered:

Replacing the Piston

1) Dismount the Syringe assembly from the instru-

ment, as described page 3-9.

2) Pull the piston out of the syringe body.
3) Reassemble the syringe and piston as described

page 6-7.

4) Remount the syringe assembly on the valve or

Tee junction.

Item

Part Number

100 µL piston kit

F4015072

250 µL piston kit

F4015073

500 µL piston kit

F4015074

1 mL piston kit

F4015075

5 mL piston kit

F4015076

10 mL piston kit

F4015077

25 mL piston kit

F4015078