Technical data appendix, Junction tubing, Hydraulics – Gilson 402 User Manual
Page 112

Appendix C-4
Technical Data
Appendix
C
Hydraulics
Typical values (all syringes)
Syringe capacity Max. accuracy Max.error imprecision
%
+ A %
RSD %
100
0.50
0.10
50
0.80
0.20
10
1.00
0.50
The same specifications are obtained in the
dispensing mode with liquid from the reservoir.
See OQ Document for procedure and calculations.
Syringe
Volume Accuracy
RSD
capacity (µL) pipetted (µL)
%
µL
%
µL
100
1
5.0
0.05
3.0
0.03
250
2.5
3.0
0.075
2.0
0.05
500
5
3.0
0.15
2.0
0.1
1000
10
3.0
0.3
1.0
0.1
5000
50
3.0
1.5
0.8
0.4
10000
100
3.0
3
0.5
0.5
25000
250
3.0
7.5
0.5
1.25
For 1 % syringe capacity, typical
values are given.
Junction Tubing
(between valve and Tee)
Material: FEP, with two PVDF union connectors.
Tubing dimensions: 2.0 mm i.d., 70 mm length.
Tubing dead volume: 220 µL.