3-10

Installation

3

Upgrading to Dual-syringe Model

Upgrading to Dual-syringe Model

The single syringe 402 can be upgraded to improve

the versatility of the system.

To upgrade to the Dual-syringe 402, you must

install a Tee-junction module (kit reference F4015010)

at the right-hand-side of the 402. You must first

install the syringe module (includes motor etc.) and

then fit the Tee-junction.

Before installing the module: make sure that the ON/

OFF switch is in the OFF position, remove the power

cord, and disconnect any hydraulic connections

from the 402.