Operation - standard use, Running a method (run) – Gilson 402 User Manual

Page 47

background image

4-11

Operation - Standard Use

4

‘Homing ...’ appears on the screen until the syringe(s)

have returned to their uppermost (home) positions,

during which time any liquid contained in the

transfer-tubing is expelled from the probe-end.

After this the Action Screen reappears.
Press ESC to return to the start-up screen for the

current file.
Whenever you see ‘Homing’ on the screen, the

syringe piston(s) are returned to their uppermost

position(s). As a result, some (or all) of the liquid

contained in the transfer-tubing will be expelled

from the probe-end. You must ensure that the

probe end of transfer-tubing is already in a

suitable waste container.

Running a Method (Run)

Introduction

The Run soft-key appears, after you have entered

your Method. Press Run to start the Method cur-

rently shown on the screen (at the first task, in this

example Dispense).
If you want to run a different program, you must

key in its file number and press ENTER, before

pressing Run.
After pressing Run, a series of in run screens

appears. During the run, you must follow the

instructions on the screen, and press the command

button on the probe when prompted to do so by

the command light on the probe.
You press the command button on the probe to

aspirate and to dispense:
- a steady light means you are going to aspirate or

dispense liquid,

- a flashing light means you are going to aspirate air.
When you have finished, or wish to change pro-

grams, press STOP, to instantly halt the program.

Pressing STOP

When you press STOP, the motion of the syringe(s)

ceases and the Abort Screen appears. Press Proceed,

when you want to continue with your Method.
Press Abort, when you want to stop. Unless the

syringe(s) are already at their uppermost (home)

positions, Initialization Screen 2 reappears.

Essential Steps

Run Screen (1)

Run Screen (33)

Abort Screen