Technical data appendix, Syringes performance specifications, Protocol – Gilson 402 User Manual
Page 111: Hydraulics

Appendix C-3
Technical Data
Appendix
C
Hydraulics
Syringes
Performance Specifications
The following values are obtained using a defined
protocol. If this protocol is followed, the obtained
values must be inferior or equal to the listed figures.
Syringe capacity (µL)
100
250
500
1000
5000
10000
25000
Maximum Flow (mL/min)
4
10
20
40
120
120
120
Minimum Specified Volume (µL)
10
25
50
100
500
1000
2500
Nominal Volume
(µL)
100
250
500
1000
5000
10000
25000
Protocol
Configuration: 1 mL tubing for volumes up to 1 mL,
tapered transfer-tubing of 2 mL (standard accessory)
for volumes up to 2 mL, 5 mL or 10 mL tapered
transfer tubings are used for volumes up to 10 mL.
Transfer of de-ionized water using the Pipette Task,
aspiration of liquid from a vial and delivery at the
same flow rate into another vial containing water.
Syringe
Volume Systematic error Random error
capacity (µL) pipetted (µL)
%
µL
%
µL
10
1.8
0.18
0.8
0.08
100
50
1.5
0.75
0.25
0.125
100
1
1
0.15
0.15
20
1.5
0.38
0.8
0.2
250
125
1.0
1.25
0.25
0.32
250
0.8
2
0.15
0.38
50
1.5
0.75
0.50
0.25
500
250
1.0
2.5
0.25
0.63
500
0.8
4
0.15
0.75
100
1.5
1.5
0.50
0.5
1000
500
0.8
4
0.25
1.3
1000
0.6
6
0.15
1.5
500
1.2
6
0.5
2.5
5000
2500
0.8
20
0.25
6.3
5000
0.6
30
0.15
7.5
1000
1.2
12
0.5
5
10000
5000
0.8
40
0.25
12.5
10000
0.5
50
0.15
15
2000
1.0
25
0.50
5
25000
12500*
0.8
100
0.25
31.3
25000*
0.5
125
0.15
37.5
* Dispensing mode only.
The values are obtained by
pipetting de-ionized water
(grade 3, ISO3696) at 21 °C ± 0.5 °C,
relative humidity > 50 % and
athmospheric pressure 1013 MPa.