Operation - advanced use, Running a method (run), Software – Gilson 402 User Manual
Page 79: In-run prompts

5-29
Operation - Advanced Use
5
Value for use with
File Number
TASK/Parameter
Dual-syringe Model
Single-syringe Model
Method 92
Dispense
Sample Source
Reservoir (RESV)
Reservoir (RESV)
Fractions
Same
Same
Nb of Fractions
3
3
Common Volume
1000
1000
Method 93
Dispense
Sample Source
Probe
Probe
Fractions
Different
Different
Nb of Fractions
3
3
Fraction 1
50
500
Fraction 2
100
700
Fraction 3
200
500
Adjust Source
Reservoir (RESV)
Reservoir (RESV)
Final Volume
1000
1000
Method 94
Pipette
Fractions
Different
Different
Nb of Fractions
4
4
Fraction 1
50
300
Fraction 2
80
400
Fraction 3
100
500
Fraction 4
200
600
Running a Method (RUN)
The Run soft-key appears, after you have entered your
Method. Press Run to start the Method currently
shown on the screen. If you want to run a different
program, you must key in its file number and press
ENTER, before pressing Run. In either case a series
of in run screens appears, starting with the in-run
prompts (if any).
In-run Prompts
Values for any parameters defined as in-run prompts
(using a question mark) must be confirmed, or
changed, before the Method runs.
After pressing Run, the software prompts you for
action: you must press ENTER to confirm the value
shown on the screen, or to change it, key in a new
value and then press ENTER.
When the Method runs you must follow the instruc-
tions on the screen. You press the command button
on the probe to aspirate and to dispense, when
prompted to do so by the command light on the probe:
- a steady light means you are to aspirate or dispense
liquid,
- a flashing light means you are to aspirate air via
the probe.
Software