3 i/o unit replacement, 1 accessing the i/o unit, 2 removing the i/o unit – FUJITSU SPARC ENTERPRISE M5000 User Manual
Page 154: I/o unit replacement, Accessing the i/o unit, Removing the i/o unit

8-10
SPARC Enterprise M4000/M5000 Servers Service Manual • August 2009
8.3
I/O Unit Replacement
The I/O units are cold replacement components.
8.3.1
Accessing the I/O Unit
Caution –
There is an electrical hazard if the power cords are not disconnected. All
power cords must be disconnected to completely remove power from the server.
●
Power off the server.
This step includes turning the key switch to the Service position, confirming that the
POWER LED is off, and disconnecting power cables. See
8.3.2
Removing the I/O Unit
Caution –
Use proper ESD grounding techniques when handling components. See
“Safety Precautions” on page 1-1
1. Disconnect and label all the cables that connect to the I/O unit.
2. Remove the cassettes from the I/O unit.
See
“Removing the PCI Cassette” on page 8-5
3. Disconnect the cabinet end of the cable management arm.