ITC Hemochron Signature+ Whole Blood Microcoagulation System User Manual
Page 47

47
Error Message
Cause
Corrective Action
MACHINE LOCKED 
CALL SUPERVISOR 
The EQC failure limit has been 
exceeded.  
Contact the Supervisor.
Sample Not Seen
Sample has not reached front 
detector in specified time period. 
Repeat test with new cuvette. If 
message persists, contact ITC 
Technical Service. 
Cuvette Removed
Cuvette was prematurely 
removed from instrument while 
testing in progress. 
Repeat test with new cuvette.
Sample Too Large/ 
Sample Too Small 
Excess or insufficient sample.
Repeat test with new cuvette.
Premature Sample
Sample was detected at front 
detector before specified time 
period. May occur if sample is 
added before pump-priming 
sequence is complete. 
Repeat test with new cuvette. If 
message persists, contact ITC 
Technical Service. 
HARDWARE FAULT
Electrical disturbance. This may 
be caused by equipment or 
devices such as centrifuges, etc. 
Repeat test using different AC 
outlet location. 
MEMORY FAULT
Malfunction in the computer's 
memory. 
Contact ITC Technical Service.
START Timed Out
START key was not pressed 
within 5 minutes after entering 
ready mode. 
Remove cuvette and repeat test 
with new cuvette. 
COD=
Cuvette cannot be identified by 
instrument. 
Remove cuvette and repeat test 
with new cuvette. 
Invalid code
Invalid correlation data entered 
by Operator. 
Re-enter information.
No Data Stored
Operator attempted to print data 
when none was stored in 
database. 
N/A
Use new cuvette
Used cuvette was reinserted into 
instrument. Occurs if cuvette 
channel is blocked or if sample 
is added before pump priming 
sequence is complete. 
Remove cuvette and repeat test 
with new cuvette. 
Access Denied
Date/Time cannot be changed if 
QC Lockout is required, or 
Date/Time function is denied 
through Configuration Manager. 
Supervisor needs to 
reconfigure instrument using 
HCM. 
Disallowed Assay
Assay performance has been 
prohibited via Configuration 
Manager. 
Supervisor needs to 
reconfigure instrument using 
HCM. 
Cannot erase DB
Erasing patient and/or QC 
database is not allowed. 
Supervisor needs to reconfigure 
instrument using HCM. 
