Specifications – ITC Hemochron Signature+ Whole Blood Microcoagulation System User Manual
Page 10

10
SPECIFICATIONS
Specifications for the HEMOCHRON Jr. Signature+ Whole Blood Microcoagulation System are listed below.
Dimensions and Weight
Depth
9 cm (3.5 in)
Width
19 cm (7.5 in)
Height
5 cm (2.0 in)
Weight
0.60 kg (1.3 lbs)
Performance
Test Precision
≤ 10% C.V.
Operation
Test Chamber
1
Timing Range
0 second to 1005 seconds
Incubation Temperature
37 °C ±1.0 °C
Incubation Warm-Up Time
30 seconds to 90 seconds
Full-Charge Operating Time
2 to 3 hours (minimum)
Battery Life
500 recharges
Battery Type
Nickel Metal Hydride
Throughput (Full Charge)
49 test cycles (at 150 sec per test)
17 test cycles (> 500 sec per test)
Operating Environment
15 °C to 30 °C
AC/DC Power Module
USA/Canada
Input
Power
120 VAC, 60 Hz, 35 VA
Output
Power
9.0 VDC, 1.38 Amps maximum
Europe
Input
Power
230 VAC, 50 Hz, 30 VA
Output
Power
9.0 VDC, 1.15 Amps maximum
Japan
Input
Power
100 VAC, 50/60 Hz, 40 VA
Output
Power
9.0 VDC, 1.3 Amps maximum
Calibration
The Signature+ instrument is calibrated at the manufacturer facility to test and verify all functions. The
instrument is also self-calibrating, as all instrument functions are continuously monitored and verified by
the instrument software when a test is performed. The instrument does not require additional calibration
by the user.