Starting the instrument – ITC Hemochron Signature+ Whole Blood Microcoagulation System User Manual
Page 14

14
The cable configurations used for connecting a serial printer or a computer are shown below:
Connecting the instrument to the Seiko
DPU-414 printer with a DB9S connector
Connecting the instrument to any personal
computer with a DB9P connector
STARTING THE INSTRUMENT
Start the instrument by pressing and holding down the START key or by inserting a cuvette:
Note: Insert the cuvette with the label facing up. Push the cuvette all the way in until it stops.
Allow cuvettes to reach room temperature (15 to 30 °C) before opening the pouch. This may
take up to 60 minutes. Consult the cuvette package insert for additional information
concerning cuvette storage and handling.
The prompts that are displayed after the instrument is started depend upon whether the START key or a
cuvette is used and whether or not lockouts are enabled (see the Configuration Manager section on page 26):
Lockout(s)
Enabled?
No
Yes
Press START
"HEMOCHRON
SIG+" is Displayed
Insert Cuvette
"Insert" and
"Cuvette" are
Displayed
Insert Cuvette
"HEMOCHRON
SIG+" is Displayed
Test Name is
Displayed
Lockout Prompts are
Displayed
Satisfy
Requirements
Initialization Prompts
Displayed
Press START to
Run Test