Check & Adjust

TPS1200

103

4.

MEAS (F1) to measure and to continue to the next screen.

Motorised instruments change auto-
matically to the other face.

Non-motorised instruments guide to
the other face.

)

The fine pointing has to be

performed manually in both faces.

5.

TOOLS Tilting-Axis Adjustment II

MEAS (F1) to measure the same target in the other face and to calculate
the tilting axis error.

)

If the error is bigger than the predefined limit, the procedure has to be
repeated. The tilting axis measurements of the current run are then
rejected and not averaged with the results from previous runs.

Step

Description

TPS12_025

180°