Integra CROMA24 User Manual
Page 64
64
Maintenance
Croma24 Service Manual
10. While lifting up the measuring tip of the micrometer, slide
the Carrier to the left of the stabilizer as shown in Figure 3-
20. Position it as close to the screw as possible and drop the
measuring tip onto the platen. Do this a couple of times to
ensure an accurate reading.
11. Move the left end of the stabilizer bracket until a reading of
0.075” +/- 0.003” is observed. Read only the RED numbers on
the micrometer gauge.
12. Tighten the screw on the left end of the stabilizer.
13. While lifting up the measuring tip of the micrometer, slide
the Carrier to the right of the stabilizer as shown in Figure 3-
20. Position it as close to the screw as possible and drop the
measuring tip onto the platen. Do this a couple of times to
ensure an accurate reading.
14. Move the right end of the stabilizer bracket until a reading of
0.075” +/- 0.003” is observed. Read only the RED numbers on
the micrometer gauge.
15. Tighten the screw on the right end of the stabilizer.
16. Repeat these steps as many times as necessary to ensure an
accurate alignment.