Specifications

Specifications

Table A-1 lists the specifications of the dc source. Specifications are warranted over the ambient
temperature range of 0 to 55

°

C. Unless otherwise noted, specifications apply when measured at

the rear terminals after a 30-minute warm-up period.

Table A-1. Performance Specifications

Parameter

HP 6631B

HP 6632B
HP 66332A

HP 6633B

HP 6634B

Output Ratings

Voltage:

Current:

0

8 V

0

10A

0

20 V

0

5A

0

50 V

0

2 A

0

100 V

0

1A

Programming Accuracy
(@ 25C

±

5

°

C)

Voltage:

+Current:

0.05% + 5 mV

0.05% + 4 mA

0.05% + 10 mV

0.05% + 2 mA

0.05% + 20 mV

0.05% + 1 mA

0.05% + 50 mV

0.05% + 0.5 mA

DC Measurement
Accuracy
(via HP-IB or front panel
meters with respect to
actual output at 25C

±

5

°

C)

Voltage:

Low Current range

20 mA to +20 mA:

High Current range

+20 mA to +rated I:

20 mA to

rated I:

0.03% + 2 mV

0.1% + 2.5

µ

A

1

0.2% + 1 mA

2

0.2% + 1.6 mA

0.03% + 3 mV

0.1% + 2.5

µ

A

1

0.2% + 0.5 mA

2

0.2% + 1.1 mA

0.03% + 6 mV

0.1% + 2.5

µ

A

1

0.2% + 0.25 mA

2

0.2% + 0.85 mA

0.03% + 12 mV

0.1% + 2.5

µ

A

1

0.2% + 0.25 mA

2

0.2% + 0.85 mA

Ripple and Noise
(in the range of 20 Hz to
20 MHz with outputs
ungrounded or with
either terminal grounded)

Normal mode voltage

(rms/p-p):

Fast mode voltage

(rms/p-p):

Current (rms):

0.3 mV/3 mV

3

1 mV/10 mV

3 mA

0.3 mV/3 mV

3

1 mV/10 mV

2 mA

0.5 mV/3 mV

3

1 mV/15 mV

2 mA

0.5 mV/3 mV

3

2 mV/25 mV

2 mA

Load Regulation

4

(change in output voltage
or current for any load
change within ratings)

Voltage:

Current:

2 mV

2 mA

2 mV

1 mA

4 mV

1 mA

5 mV

1 mA

Line Regulation
(change in output voltage
or current for any line
change within ratings)

Voltage:

Current:

0.5 mV

1 mA

0.5 mV

0.5 mA

1 mV

0.25 mA

1 mV

0.25 mA

Transient Response
Time

4

Normal mode:

Fast mode:

< 100

µ

s

< 50

µ

s

(for the output voltage to recover to its previous level within 0.1% of the voltage rating of the unit or 20 mV following a
change in load current of up to 50% of the output current rating)

1

This specification may degrade slightly when the unit is subjected to an RF field

3 V/meter.

2

Applies in SCPI mode, with current detector set to DC. With current detector set to ACDC, accuracy is 0.2% + four

times the fixed error value. In COMPatibility mode, accuracy is 0.2% + six times the fixed error value.

3

For HP 6631B, 6632B, and 66332A (from 1 MHz to 20 MHz ) = 0.3mV/15mV;

For HP 6633B and 6634B (from 1 MHz to 20 MHz ) = 0.5mV/15mV

4

Applies at rear terminals with unit set to remote sensing and with sense terminals externally jumpered to their

respective output terminals.