When the environmental conditions preclude the “3P” measurement mode, as for example 
in a city center. In TT plants it is possible to measure the earth resistance with the 
simplified “2P” method that gives a higher value than the “3P” mode does. 
An auxiliary rod is necessary; it must have an insignificant earth resistance and must be 
independent of the earth plant under test. 
In the picture above, a water tube is used as the auxiliary rod. But, any metallic object 
fixed to the ground could replace it. 
 
Example: if the operator takes three consecutive measurements, the instrument displays 
1
st
measurement:
main display = measured resistance value (Ex: 0.90
Ω)
secondary display on the left-hand side = 001 (no. of measurements = 1 
means that 1 earth measurement has been taken) 
secondary display = average of the measurements taken (in case just one 
measurement has been taken the medium value is equal to the measured 
value, in this case 0.90
Ω)
2
nd
measurement:
main display = measured resistance value (Ex: 0.96
Ω)
secondary display = average of the measurements taken ((Val1+Val2)/no. 
of measurements = (0.90+0.96)/2 = 0.93
Ω)
3
rd
measurement:
main display = measured resistance value (Ex: 0.93
Ω)
secondary display = average of the measurements taken ((Val1+Val2)/no. 
of measurements = (0.90+0.96+0.93)/3 = 0.93
Ω)
 
 
Note: Any test with a result over 1999
Ω is not inserted in the calculation of medium
value.
 
 
Example: 1
st
measurement:
main display: 1.07
Ω
secondary display: 1.07
Ω
2
nd
measurement
main display: 4.15
Ω
secondary display: 2.61
Ω
3
rd
measurement (not inserted in the medium value)
main display: >1999
Ω
secondary display: 2.61
Ω