Amprobe, Technical specifications, Dmiii multitest – Amprobe DMIII-Multitest Power-Quality-Recorder User Manual
Page 78: En-76
AMPROBE
DMIII MULTITEST
EN-76
14. TECHNICAL SPECIFICATIONS
14.1. TECHNICAL FEATURES
Accuracy is indicated as [% of reading + number of digits]. It refers to the following
atmospheric conditions: a temperature of 23°C ± 5°C with a relative humidity < 60%.
14.1.1. Safety Test functions
LOW
: 200mA CONTINUITY TEST (AUTO, RT+, RT- MODE)
Range [
]
Resolution [
]
Accuracy(*)
0.01
9.99
10.0
99.9
0.01
0.1
(2% Reading + 2 digit)
(*) After Test leads calibration
Test Current
> 200mA DC per R
5
(Test leads included)
Resolution for Test current:
1mA
Open Circuit Voltage
4V
V0
24V
INSULATION TEST
Test Voltage [V]
Range [M
]
Resolution [M
]
Accuracy
50
0.01
9.99
0.01
(2% Reading + 2 digit)
10.0
49.9
0.1
50.0
99.9
0.1
(5% Reading + 2 digit)
100
0.01
9.99
0.01
(2% Reading + 2 digit)
10.0
99.9
0.1
100.0
199.9
0.1
(5% Reading + 2 digit)
250
0.01
9.99
0.01
(2% Reading + 2 digit)
10.0
199.9
0.1
200
249
1
250
499
1
(5% Reading + 2 digit)
500
0.01
9.99
0.01
(2% Reading + 2 digit)
10.0
199.9
0.1
200
499
1
500
999
1
(5% Reading + 2 digit)
1000
0.01
9.99
0.01
(2% Reading + 2 digit)
10.0
199.9
0.1
200
999
1
1000
1999
1
(5% Reading + 2 digit)
Open circuit Test Voltage
<1.3 x Nominal Test Voltage
Short Circuit Current
<6.0mA with 500V Test Voltage
Nominal Test Current
500V
>2.2mA with 230k
other
>1mA with 1k
*Vnom
FREQUENCY MEASUREMENT
Range [Hz]
Resolution [Hz]
Accuracy
47.0
63.6
0.1
(0.1%Reading+1 digit)
RCD and LOOP function are active only for 50Hz
0,5Hz frequency
PHASE ROTATION : VOLTAGE MEASUREMENT
Range [V]
Resolution [V]
Accuracy
0
460V
1
(3%Reading + 2digit)
GROUND TEST: RESISTANCE MEASUREMENT WITH EARTH RODS
Range RE [
]
Resolution [
]
Accuracy
0.01
– 19.99
20.0
– 199.9
200 - 1999
0.01
0.1
1
(5% Reading + 3 digit)
Test Current
<10mA
– 77.5Hz
Open circuit Test Voltage
<20V RMS
GROUND TEST: RESISTIVITY MEASUREMENT
Range
Resolution
Accuracy
0.60
19.99
m
20.0
199.9
m
200
1999
m
2.00
99.99k
m
100.0
125.6k
m(*)
0.01
m
0.1
m
1
m
0.01 k
m
0.1 k
m
(5% Reading + 3 digit)
(*) setting distance = 10m
Test Current
<10mA
– 77.5Hz
Open circuit Test Voltage
<20V RMS