Amprobe – Amprobe DMIII-Multitest Power-Quality-Recorder User Manual
Page 36

AMPROBE
DMIII MULTITEST
EN-34
6.4.3. "2-W", "3-W" and "
" faulty cases
If the instrument detects
the
External
Power
supply
adapter
connected to instrument
will show the message
displayed to the right.
EARTH
05.06.01
-
- -
Vd= ---V
Test:04
RAVG=0.74
REMOVE POWER
3-W
Disconnect the External
Power Supply Adapter
FUNC CLR
If the Instrument detect
a voltage values higher
than 5V the instrument
will shows the screen
displayed to the right.
EARTH
05.06.01
Warning
symbol:
Voltage
noise in input.
-
- -
Vd= 230V
Test:04
RAVG=0.74
VOLT IN INPUT
3-W
FUNC CLR
The
message
"Rc
high" indicates that the
instrument
can't
produce the minimum
current necessary for
measurement.
Check
that the terminals are
correctly
connected
and the Auxiliary earth
rod connected to T4
(yellow conductor) has
not been inserted in a
pebbly
or
poorly
conductive ground. If
necessary pour some
water around the rod.
EARTH
05.06.01
-
- -
Vd= 1V
Test:04
RAVG=0.74
Rc HIGH
3-W
Message
"Rc":
Check
Auxiliary Earth rods.
FUNC CLR