Amprobe – Amprobe DMIII-Multitest Power-Quality-Recorder User Manual
Page 26

AMPROBE
DMIII MULTITEST
EN-24
6.2.3. Results of "TMR" mode
At the end of the test if
the
Insulation
resistance
is
lower
than R
MAX
(see Table2)
and the instrument
generated
the
Nominal test Voltage,
the instrument emits a
double
beep
indicating the positive
outcome of the test
and displays a screen
similar to the screen to
the right.
M
05.06.01
Insulation Resistance
1.07
M
514V 15s
TMR 500V TIME:60s
Voltage during the Test
Duration of the Test
Test mode
FUNC VNOM
The displayed result can be stored by pressing the SAVE key twice (according to
paragraph 9.1).
If
the
Insulation
resistance
is
higher
than R
MAX
(see Table2),
the instrument emits a
double beep at the end
of the test indicating
the positive outcome
of the test and displays
a screen similar to the
screen to the right.
M
05.06.01
Maximum resistance value
that can be measured
(999
is displayed if a
rated voltage of 500V was
selected, see Table2).
> 999
M
523V 15s
TMR 500V TIME:60s
The symbol ">" means that
the resistance value is
higher than R
MAX
.
Test duration
FUNC VNOM
The displayed result can be stored pressing the SAVE key twice (according to
paragraph 9.1).