5.5.
RESET
This option re-establishes the default settings of the instrument.
ANALYZER CONFIG:
Type of electrical equipment:
Three Phase 4 wires
Frequency:
not modified
Clamp full scale:
not modified
Clamp type:
not modified
Transforming ratio of voltmetric transformers:
1
Password:
disabled
RECORDER CONFIG:
Start:
manual (the recording is started
at 00 sec mark on clock after pressing
START/STOP)
Stop:
manual
Integration period:
15min
Recording of harmonics:
ON
Recording of voltage anomalies (sags and surges):
ON
Voltage reference for detection of sags and surges:
110V
Upper limit for detection of sags and surges:
6%
Lower limit for detection of sags and surges:
10%
Selected voltages:
V1, V2, V3
Selected voltage harmonics:
THD, 01, 03, 05, 07
Selected currents:
I1, I2, I3, IN
Selected current harmonics:
THD, 01, 03, 05, 07
CO-GENERATION:
OFF
Selected powers, Pf and cos
:
Pt, P1, P2, P3
Qti, Q1i, Q2i, Q3i
Qtc, Q1c, Q2c, Q3c
St, S1, S2, S3
Pft, Pf1, Pf2, Pf3
dpft, dpf1, dpf2, dpf3
Energies:
Eat, Ea1, Ea2, Ea3
Erit, Eri1, Eri2, Eri3
Erct, Erc1, Erc2, Erc3
Est, Es1, Es2, Es3
The RESET command will not erase the instrument’s memory.