Amprobe – Amprobe DMIII-Multitest Power-Quality-Recorder User Manual
Page 21

AMPROBE
DMIII MULTITEST
EN-19
6.1.5. "AUTO", RT+", "RT-" faulty cases
If the instrument detect
the
External
Power
supply
adapter
connected to instrument
will show the message
displayed to the right.
LOW
05.06.01
-
.- -
R+ R-
---
---
---mA ---mA
REMOVE POWER
AUTO 0.11
Disconnect the External
Power Supply Adapter
FUNC CAL
If the terminal voltage is
higher than 15V, the
instrument
does
not
perform the test and
displays the screen to
the right for 5 seconds.
LOW
05.06.01
-.- -
R+ R-
-.--
-.--
---mA ---mA
VOLT IN INPUT
AUTO 0.11
ATTENTION: the test was
not performed because of
voltage at the terminal
ends.
FUNC CAL
In the case that:
R
CALIBRATION
>R
MEASURED
the instrument displays
the screen to the right.
LOW
05.06.01
0.00
R+ R-
0.00
0.00
219mA 219mA
CAL > RES
AUTO 0.11
ATTENTION:
R
CALIBRATION
>R
MEASURED
FUNC CAL