6.1.5. "AUTO", RT+", "RT-" faulty cases 
 
 If the instrument detect
the
External
Power
supply
adapter
connected to instrument 
will show the message 
displayed to the right. 
LOW
05.06.01
-
.- -
R+ R-
---
---
---mA ---mA
REMOVE POWER
 
AUTO 0.11
Disconnect the External 
Power Supply Adapter
FUNC CAL
 
 
 If the terminal voltage is
higher than 15V, the 
instrument 
does
not
perform the test and 
displays the screen to 
the right for 5 seconds. 
LOW
05.06.01
-.- -
R+ R-
-.--
-.--
---mA ---mA
VOLT IN INPUT
AUTO 0.11
ATTENTION: the test was 
not performed because of 
voltage at the terminal 
ends.
FUNC CAL
 
 
 In the case that:
R
CALIBRATION
>R
MEASURED
 the instrument displays 
the screen to the right. 
LOW
05.06.01
0.00
R+ R-
0.00
0.00
219mA 219mA
CAL > RES
AUTO 0.11
ATTENTION: 
R
CALIBRATION
>R
MEASURED
FUNC CAL