Dmiiis, Frequency measurement – Amprobe DM-111S Current-Data-Logger User Manual
Page 50
DMIIIS
EN - 48
13.1.6. Measurement of harmonics
Voltage
Range
Accuracy
Resolution
DC
– 25h
26h
– 33h
34h
– 49h
(5.0%+2digit)
(10.0%+2digit)
(15.0%+2digit)
0.1V
The voltage harmonics will be null under the following threshold:
-
DC: if <1V or <2% of 1
st
harmonic
-
1
st
harmonic: if <2V
-
2
nd
49
th
: if <1V or <2% 1
st
harmonic
Current
Range
Accuracy
Resolution
DC
– 25h
26h
– 33h
34h
– 49h
(5.0%+2digit)
(10.0%+2digit)
(15.0%+2digit)
0.1A
The current harmonics will be null under the following threshold:
-
DC: if <2% of 1
st
harmonic or < 0,2% of clamp full scale
-
1
st
harmonic: if < 0,2% of clamp full scale
-
2
nd
49
th
: if<2% 1
st
harmonic or < 0,2% of clamp full scale
Setting the FLEX option the DC component will be ignored.
13.1.7. Frequency measurement
Instrument set to 50Hz
Range
Resolution
Accuracy
47
53
0,1Hz
(1.0%+1digit)
Instrument set to 60Hz
Range
Resolution
Accuracy
57
63,6
0,1Hz
(1.0%+1digit)
13.1.8. Compliance
DM-IIIS complies with the standards prescribed for:
class 2 EN61036
– static counters of active energy
class 3 IEC1268
– static counters of reactive energy
13.1.9. Temperature drift
Temperature drift: 0,1 x accuracy/K
13.1.10.
Safety
The instrument complies to the standards:
EN 61010-1 + A2(1996)
Insulation:
Class 2
Pollution:
2
Over-voltage category:
CAT III 370V~ (Phase-Earth)
CAT III 635V~ (Phase-Phase)