Specifications, Technical parameters – RIGOL DS6000 Series User Manual
Page 60

RIGOL
16
RP1003C/RP1004C/RP1005C User’s Guide
Specifications
The following parameters are applicable when the instrument has
been working for more than 30 minutes under the specified
operating temperature (10℃ to 30℃) and humidity (0%RH to
80%RH).
Technical Parameters
RP1003C:
Bandwidth
DC to 50MHz (-3dB), refer to Appendix 1
Amplitude Characteristics (RP1003C)
Rise Time
≤7ns
Linear Response Range of
Input Current
30Arms, refer to Appendix 2 Relation
between Max Input Current and
Frequency (RP1003C)
Maximum Peak Current
50A peak, non-continuous
Gain
0.1V/A
Gain Accuracy
±1.0%rdg±1mV, ≤30A
±2.0%rdg, 30A to 50A peak
non-continuous (DC and 45Hz to 66Hz,
within the linear response range of the input
current)
Noise
≤2.5mArms (for 20MHz band measuring
instrument)
Input Impedance
Refer to Appendix 3 Input Impedance
(Typical) (RP1003C)
Gain Accuracy Drift
≤±2% (0℃ to 40℃ temperature, input of
50Hz, 30A)