RP1003C/RP1004C/RP1005C User’s Guide
9
Note: When disconnecting the output connector and the
measuring instrument, be sure to release the lock by rotating
the collar in the opposite direction of the arrow as shown in
the figure above before pulling off the connector. Forcibly
pulling the connector without releasing the lock or pulling on
the cable can damage the terminator.
4. Make sure the current sensor is locked (for RP1003C and
RP1004C, the slide switch should be at the LOCK position; for
RP1005C, LOCK should be displayed on the slide switch and
UNLOCK should disappear.
5. Turn on the power adaptor and press the demagnetizing
switch on the terminator. The demagnetizing process takes
about one second. During the demagnetizing, the current
probe outputs a demagnetizing waveform.
Note: Do not demagnetize the current probe when it is
clamping a conductor to be measured. Demagnetizing causes
current to flow into the conductor, which may damage the
parts in the circuit to be measured.
RP1003C and RP1004C:
RP1005C:
Ч
Ч
Demagnetizing
Demagnetizing