RIGOL DS6000 Series User Manual
Page 40

RIGOL
VI
RP1003C/RP1004C/RP1005C User’s Guide
Make sure that the measuring instrument connected to the
output terminal of the current probe is equipped with a
protective earthing with double-insulation construction.
If the measuring instrument connected to the output terminal
(BNC) of the current probe is equipped with any other
measurement terminals, take the following precautions to
ensure that the other instrument does not form a bridge
between the probe and any hazardous live of a part.
a) Isolate the terminal of the probe from other terminals on
the measuring instrument using basic insulation
conforming to the requirements of measurement
category, working voltage and pollution degree.
b) If basic insulation requirement cannot be met between
the terminal to which the current probe is connected and
other terminals on the measuring instrument, make sure
that the voltage input to the measurement terminal does
not exceed the safe voltage level.
When using a measuring instrument that does not provide
isolation between its input terminals and chassis or other
input terminals, please do not connect the ground terminal to
any non-ground potential. Otherwise, short-circuit current will
flow through the power adaptor or current probe from the
ground terminal, which could cause an electrical accident or
damage.
RP1003C and RP1004C: