POWER
OPTIONS
85-265 VAC 50/60 Hz, 90-265 VDC, 20 W max
or jumper selectable 12/24 VDC
± 10%, 15 W max
FUSE
Required external fuse: UL Recognized, 5 A max, slow
blow; up to 6 meters may share one 5 A fuse
ISOLATION
500 V AC/DC potential allowed between voltage and current
input channels. Channels isolated by 3 M
Ω impedance
4 kV input/output-to-power line
500 V input-to-output or output-to-P+ supply
OVERVOLTAGE
CATEGORY
Installation Overvoltage Category II:
Local level with smaller transient overvoltages than
Installation Overvoltage Category III.
ENVIRONMENTAL Operating temperature range: -40 to 65°C
Storage temperature range: -40 to 85°C
Relative humidity: 0 to 90% non-condensing
CONNECTIONS
Removable screw terminal blocks accept 16 to 28 AWG
wire, RJ45 for external relays, digital I/O, and serial
communication adapters.
ENCLOSURE
1/8 DIN, high impact plastic, UL 94V-0, color: black
MOUNTING
1/8 DIN panel cutout required:
3.622" x 1.772" (92 mm x 45 mm)
Two panel mounting bracket assemblies are provided.
TIGHTENING
TORQUE
Screw terminal connectors: 5 lb-in (0.56 Nm)
OVERALL
DIMENSIONS
4.68" x 2.45" x 5.64" (119 mm x 62 mm x 143 mm)
(W x H x D)
WEIGHT
9.5 oz (269 g)
WARRANTY
3 years parts & labor
Voltage and Current Inputs
HIGH VOLTAGE
INPUT
One high voltage input (Channel V)
0-300 VDC or VAC; Switch Selectable, Modbus PV (Slave)
HIGH CURRENT
INPUT
One high current input (Channel A)
0-5 AAC or ADC; Switch Selectable, Modbus PV (Slave)
CHANNELS
Channel A, Channel V, Channel P (Math Channel for
Apparent Power)
AC/DC
SELECTION
Switch selectable for AC or DC inputs. Channels A and V
share AC/DC selection.
APPARENT
POWER MATH
Apparent power P calculated as P = ((A * V) + C) * F
PROGRAMMA-
BLE CONSTANTS
Constant C (Adder): -99.999 to 999.999, default: 0.000
Constant F (Factor): 0.001 to 999.999, default: 1.000