beautypg.com

Error message (error), Minimum input span – Precision Digital PD6400 User Manual

Page 78

background image

Model PD6400 High Voltage & Current Meter

Instruction Manual

78

Error Message (Error)

An error message indicates that the calibration or scaling process was
not successful.
After the error message is displayed, the meter reverts to input 2 during
calibration or scaling and to input 1 during internal calibration, allowing
the appropriate input signal to be applied or programmed.

The error message might be caused by any of the following conditions:
1. Input signal is not connected to the proper terminals, or it is

connected backwards.

2. Wrong signal selection in Setup menu.
3. Minimum input span requirements not maintained.

Minimum Input Span

The minimum input span is the minimum difference between input 1 and
input 2 signals required to complete the calibration or scaling of the meter.

Input Range

Input 1 & Input 2 Span

+/- 0-5 ADC

+/- 0.005 ADC

0-5 AAC

0.010 AAC

+/- 0-300 VDC

+/- 0.1 VDC

0-300 VAC

0.3 VAC