Model PD6400 High Voltage & Current Meter

Instruction Manual

54

Signal Loss or Loop Break Relay Operation

The following graph shows the loop break relay operation for a high
alarm relay.


When the meter detects a break in the 4-20 mA loop, the relay will go to
one of the following selected actions:

1. Turn On (Go to alarm condition)
2. Turn Off (Go to non-alarm condition)
3. Ignore (Processed as a low signal condition)

Relay

LED

Input

Loop Break = Ignore

de-energized

energized

off

on

Relay

LED

Loop Break = Off

Relay

LED

Loop Break = On

Set

Reset