beautypg.com

Precision Digital PD6400 User Manual

Page 77

background image

Model PD6400 High Voltage & Current Meter

Instruction Manual

77

8. The display moves to the high input calibration (C Hi). Apply the

high input signal and press Enter.

9. Set the display for the high input calibration, in the same way as it

was set for the low input calibration (e.g. 300 VDC).

The graphic above shows the calibration of the current input. The
voltage input is calibrated in a similar way.

Tips:

Low and high input signals can be any valid values within the
range of the meter.

Observe minimum input span requirements between input 1
and input 2.

Low input should be less than high input signal.