Operating Conditions

MKT-7.5.3-L-183 | Rev 0

The Digital ZipSpin centrifuge is designed to be safe under the following conditions when operated with necessary
precautions:

• For indoor use only.
• Ambient temperature of 4°C to 65°C.
• Main supply voltage fluctuations not to exceed +/- 10% of the nominal voltage.
• Position at least 20 cm from walls, 20 cm from ceiling and 20 cm between other products.
• Make sure to always install the centrifuge on stable flat surfaces only & place it near the electrical outlet.
• Make sure to always load the rotor symmetrically every time. Each tube should be counterbalanced by another tube.
• Make sure to always use only specified micro centrifuge tubes made from plastic and designed to withstand

centrifugal forces of at least 2000 g.
• Do not use the centrifuge in any manner not specified in these instructions.
• Do not operate the centrifuge without the rotor being properly attached to the shaft. Re-check by trying to pull it up
manually, especially after rotor assembly.
• Do not move the centrifuge while the rotor is spinning.
• Do not fill tubes while they are in the rotor, liquid spillage may harm the unit.
• Do not put hands in the rotor unless the rotor has completely stopped.
• Do not centrifuge flammable, explosive or corrosive materials.
• Do not use solvents or flammables near this or other electrical equipment.
• In case of the unit dropping from the table please check the casing for any cracks/damage.
• The centrifuge and its components are non-autoclavable.
• Speeds for tube rotor are accurate to +/- 5%
• Speeds for PCR rotor are accurate to +/- 12%

RCF CHART

RPM

ROUND ROTOR

PCR STRIP

1000

2000

3000

4000

4500

5000
6000

35.5

142

319

567

718

886

1276

0.2ml

54

217

488

868

1098

1356
1952

0.2ml

47

188

424

753

953

1177
1695

0.5ml

52

208

468

832

1053

1300
1872

1.5ml

52

208

467

830

1050

1297
1868

0.2ml

Fluid Recommendations:

Whole Blood: 6,000 rpm

10 mins.

Urine:

2,800 rpm