Operating parameters – HEIDENHAIN ND 730 v.3 User Manual
Page 31

32
P41.1 to P41.3 Linear axis error compensation
Linear axis error compensation is activated via parameters
40.1 to 40.3.
Input range
[µm]
:
– 99999 to + 99999
Default setting:
0
Example: Displayed length
Ld = 620.000 mm
Actual length (as determined for example with
the VM 101 from HEIDENHAIN)
La = 619.876 mm
Difference
∆L = La – Ld = –124 µm
Compensation factor k:
k =
∆L/Ld = –124 µm/0.62 m = –200 [µm/m]
P42.1 to P42.3 Backlash compensation
Input range (mm):
+9.999 to -9.999
Default setting:
0.000 = no backlash
compensation
A change in direction might cause a clearance between
rotary encoder and table. This is referred to as backlash.
Positive backlash: The rotary encoder is ahead of the table.
The distance traversed by the table is too short (entry of
positive value). Negative backlash: The rotary encoder
follows the table. The distance traversed by the table is too
large (entry of negative value).
P43.1 to P43.3 Reference marks
One reference mark
SINGLE REF.
Distance-coded with 500 • SP
500
SP
Distance-coded with 1000 • SP
1000
SP
Distance-coded with 2000 • SP
2000
SP
Distance-coded with 5000 • SP
5000
SP
(SP = signal period)
P44.1 to P44.3 Reference mark evaluation
Evaluation
REF. X ON
No evaluation
REF. X OFF
P45.1 to P45.3 Encoder monitoring
Amplitude and frequency
monitoring
ALARM ON
No monitoring
ALARM OFF
P48.1 to P48.3 Activate axis display
Axis display active
AXIS ON
Axis display not active
AXIS OFF
P80 Function of CL key
Set to zero with CL
CL...RESET
No set to zero with CL
CL......OFF
Operating Parameters
730_770t4.pm6
21.07.2004, 08:38
32