32
P41.1 to P41.3 Linear axis error compensation
Linear axis error compensation is activated via parameters
40.1 to 40.3.
Input range
[µm]
:
– 99999 to + 99999
Default setting:
0
Example: Displayed length
La = 620.000 mm
Actual length (as determined for example with
the VM 101 from HEIDENHAIN)
La = 619.876 mm
Difference
∆
L = La – Ld = –124 µm
Compensation factor k:
k =
∆
L/Ld = –124 µm/0.62 m = –200 [µm/m]
P43.1 to P43.3 Reference marks
One reference mark
SINGLE REF.
Distance-coded with 500 • SP
500
SP
Distance-coded with 1000 • SP
1000
SP
Distance-coded with 2000 • SP
2000
SP
Distance-coded with 5000 • SP
5000
SP
(SP = signal period)
P44.1 to P44.3 Reference mark evaluation
Evaluation active
REF. X ON
Evaluation not active
REF. X OFF
P45.1 to P45.3 Encoder monitoring
Amplitude and frequency
monitoring active
ALARM ON
Amplitude and frequency
monitoring not active
ALARM OFF
P48.1 to P48.3 Activate axis display
Axis display active
AXIS ON
Axis display not active
AXIS OFF
P80 Function of CL key
Reset with CL
CL...RESET
No reset with CL
CL......OFF
P98 Dialog language
1)
German
LANGUAGE
D
English
LANGUAGE
GB
French
LANGUAGE
F
Italian
LANGUAGE
I
Dutch
LANGUAGE
NL
Spanish
LANGUAGE
E
Danish
LANGUAGE
DK
Swedish
LANGUAGE
S
Finnish
LANGUAGE
FI
Czech
LANGUAGE
CZ
Polish
LANGUAGE
PL
Hungarian
LANGUAGE
H
Portuguese
LANGUAGE
P
Operating Parameters
1)
User parameter