Page 41

6.1.10

Dry Contact Functions

Normal:

Pointer follows contact closure.
Open: zero
±3.0% of full scale.
Closed: full scale
±0.5%.

On Time:

Range: 0 to 100 %.

Accuracy:

±2% of full scale from 10% to 100%.
Zero offset:
±3.0% of full scale max.

Rate:

Range: 30 to 500 cycles per minute.

Accuracy:

±3% of scale length.

Peak Hold:

Pointer will deflect to full scale and
remain there with a 2.0 millisecond or
longer contact closure.

Allowable Contact
Resistance:

0 to 100 ohms.

Test Voltage:

7.5 to 15 VDC. The VOM jack is positive
with respect to the COM jack.

Test Current:

830 uA max.

Overload:

Continuous: 50 VRMS.
Ten seconds: 120 VRMS.

This manual is related to the following products: