6.1.5
AC Voltage
Ranges:
1.5, 3, 15, 150, 300, and 600 VAC.
Accuracy:
±3.0% of full scale 0 to 50 C.
±3.5% of full scale -40 to 85 C.
Sensitivity:
288 ohms per volt.
Overload:
Continuous: 20 VRMS or 5 times full
scale, whichever is greater, not to
exceed 700 VRMS.
Ten seconds: 30 VRMS or 10 times full
scale, whichever is greater, not to
exceed 700 VRMS.
6.1.6
AC Amps
Ranges:
1.5, 6, and 30 Amps AC.
Accuracy:
±1.5% of full scale 0 to 50 C.
±2.0% of full scale -40 to 85 C.
Burden Volts:
60 mV nominal at full scale. Leads add
34 milliohms typical resistance. Fuse in
1.5 Amp range adds 30 milliohms typical
resistance.
Overload:
For the 1.5 Amp range, the fuse blows
before damage. The 6 Amp and 30 Amp
ranges are unfused but can withstand a
50 amp overload for 30 seconds.