III. USB Communications Parameters

A. Baud Rate 230400

B. No Parity

C. 1 Start Bit 1 Stop Bit

IV. Serial Commands

1. MODEL CODE

Write Command: NA

Read Command: 00

Interpret: 9625 returned. (This is an internal code to ensure the TC-720 software is communicating with an actual

TC-720.)

2. INPUT1 (reads the value sensed by the primary sensor)

Write Command: NA

Read Command: 01

Interpret: Convert the returned hexadecimal value to decimal, and then divide by 100

10

.

3. POWER OUTPUT

Write Command: NA

Read Command: 02

Interpret: Convert the returned hexadecimal value to decimal. 511

10

represent 100% output (heating) and -511

10

represent -100% output (cooling)

4. INPUT2 (reads the value sensed by the secondary sensor)

Write Command: NA

Read Command: 04

Interpret: Convert the returned hexadecimal value to decimal, and then divide by 100

10

.

5. ALARM STATUS

Write Command: NA

Read Command: 03

Interpret:

Convert the hexadecimal number to binary and interpret bits as follows:

All zeros == means no alarms.

Bit 0==1 means HIGH ALARM1.

Bit 1==1 means LOW ALARM1.

Bit 2==1 means HIGH ALARM2.

Bit 3==1 means LOW ALARM2.

Bit 4==1 means OPEN INPUT1.

Bit 5==1 means OPEN INPUT2.

Bit 6==1 means LOW INPUT VOLTAGE

Bit 7==1 means key was pressed to store value

Bit 8==1 means OVER CURRENT DETECTED

For example, suppose the returned value is 0009

16

. This converts to 001001

2

, which means that there is a

low alarm condition for the secondary sensor and a high alarm condition for the control sensor. “Bit 0” is

the rightmost bit, or the least significant bit.

6. FIXED DESIRED CONTROL SETTING

Write Command: 1c

Read Command: 50

Interpret: To send a set temperature, multiply the decimal value by 100

10

and convert to hexadecimal. To read

the set temperature, convert the returned hexadecimal value to decimal, and then divide by 100

10

.